    Brand Name : LABNOVATION
    Model Number : LD-560
    Certification : IFCC, NGSP, CE
    Price : USD7,000.00-12,000.00/Unit
    Supply Ability : 100 Units/Day
    Delivery Time : 15-20 working days
    • Product Details
    • Company Profile

    HPLC Principle Fully Automated HbA1c Analyzer For Glycosylated Hemoglobin Diaglosis

    Fully Automated HbA1c Analyzer For Glycosylated Hemoglobin Diaglosis


    Intended use


    The Fully Automated HbA1c Analyzer is used in conjunction with suitable reagents for the quantitative detection of glycosylated hemoglobin (HbA1c) in human blood.The Fully automatedHbA1c Analyzer LD-560 offers the fast throughput of HbAlc results in 90 seconds, with Hb variant detection, providing the outstanding solution for quick and reliable diabeticmonitoring. No sample preparation and very little hands-on time by the operator is required for the LD-560 Analyzer.
    Principle
    Using high performance liquid chromatography (HPLC), the stationary phase used is a weakly acidiccation exchanger. Glycosylation of hemoglobin will result in the loss of cations on the surface of hemoglobin without being charged, and non glycosylated hemoglobin has a positive charge. According to the different charged properties of hemoglobin, eluents with different pH values and ionic strengths are used sequentially, and several hemoglobin components including HbA1c are separated by high performance liquid chromatography. Each component is detected by a single wavelength colorimeter, and the detected data is processed by a computer to calculate the percentage content of HbA1c in the total hemoglobin.
    Performance

    *HPLC, CV<2% - Accurate and Stable

    *Full Automatic smart System
    *Quality & Performance - IFCC & NGSP Certificate
    *User Friendly

    Specification


    InstrumentManufacturerLanovation
    Manufacture locationShenzhen, China
    ModelLD-560
    Dection informationPrinciple methodHPLC
    Detection methodDual wavelengh colorimetry
    Wavelength420nm
    Test Speed90 Seconds/test
    Test Speed throughtput40 Samples per hour
    Sample informationSample typesWhole blood or Pre-diluent sample
    Sample volume5μL in whole blood mode
    150μL in pre-diluted mode
    Sample capacity20 samples
    Sample loading methodContinuous automatic sample loading
    MixerNo, but probe mixing samples
    Reagent and other consumbleColumn test quantity3000T
    Reagent pack600T(Reagent)/ 3000T(Column) separate
    Eluent2
    Operation and character informationOperation screen10 inch Full color LCD display
    Data storage5000
    Built-in Thermal PrinterYes
    Performance informationCV<2%
    Test range3.0%~18.0%
     
    Package Information
    Dimensions378(L) ×380(W)×510(H)mm
    Weight28kg

    Product Composition

    Main Parts

    The analyzer consists of an auto sample injection module, a liquid chromatography separation module, a colorimetric detection and control module, a barcode scanning module, and a display printing module.



    Reagents and Consumables

    The analyzer is recommended to be used in conjunction with the Glycosylated Hemoglobin Test Reagent Kit (high performance liquid chromatography (HPLC method)) produced by our company, which includes the following components:
    Eluent A/Eluent B/Eluent C(Variant Mode)Adjusts the pH of the solution for the elution of glycosylated hemoglobin and non glycated hemoglobin.
    Hemolysis ReagentUsed to dissolve red blood cells and release hemoglobin.
    Chromatography ColumnUsed for separation and elution of glycosylated hemoglobin and non glycated.
    Hemoglobin FilterUsed to initially filter the sample after hemolysis.
    Paper RollsPrints a report of the test results.
    HPLC Principle Fully Automated HbA1c Analyzer For Glycosylated Hemoglobin Diaglosis

    Certificate


    HPLC Principle Fully Automated HbA1c Analyzer For Glycosylated Hemoglobin Diaglosis

    Quality HPLC Principle Fully Automated HbA1c Analyzer For Glycosylated Hemoglobin Diaglosis for sale
    Inquiry Cart 0
    Send your message to this supplier
     
    *From:
    *To: Labnovation Technologies, Inc.
    *Subject:
    *Message:
    Characters Remaining: (0/3000)